Item known as dry well temperature reference; wgt: 30 lbs; 12.5 depth; set point accuracy +/- 2 degrees c; setpoint stability: +/- 0.02 degrees c/day; display resolution: +/- 0.01 degrees c or f; display type: led, centrigade or fahrenheit user selectable; typicalwarm up time: 6 min 25 to 80 degrees c, 14min: 25 to 140 degrees c; typical cool down times: 10 min: 25 to -20 c; fault protection: sensor burnout protection, over temperature thermal cutout, electrical fuse: 115v, 3a sb; 2030 v, 1.5a sb; well dimensions: 2 reference wells at 6.7 mm (0.265") dia. X 155mm (6.1")deep 6 calibration wells at 9mm (0.354") dia. X 155mm (6.1") deep; calibration well capacity: 18 or 24 depending on wire size
NSN 5855-01-469-2798
NSN 4140-01-332-1973
NSN 5835-01-595-1131
NSN 5975-00-430-1220
NSN 5985-01-213-9556
NSN 5998-01-476-5369
NSN 5920-01-421-4930
NSN 5996-01-649-2142
NSN 5930-00-039-9773
NSN 5999-00-732-9109
NSN 5999-00-293-4326
NSN 6105-01-440-8795
Last Updated: 07/05/2025