Shall be capable of connecting toand testing movement thermo-couples; meter-movement resistance temperature detectors (rtds); servo thermo-couples ; and commercial thermo-couples type k/t/j and tf-30, case shall be no more than 20 inches long, 15 inches wide, and 15 inches high, case will have a handle attached and a lockingmechanism p rovided, weight shall be no more than 35 pounds; interface shall be providedthat will display values being measured or sourced while operating between 50 and 100 degrees farenheit, temperature calibrator shall have an on/off power switch; battery charge indicator; source/measure switch; output voltage tuning ability from 0-35volts down to .5mv of the desired output; mode select capability; ability to test single or double rtd indicators; ability to source 22.5 or 28.5 volts plus or minus .5mv; ability to simulate rtd and synchro indicator voltage ranges, will include 115/230v AC at 60hz power cable
NSN 5945-01-054-3722
NSN 5930-01-023-8144
NSN 6220-01-384-2253
NSN 6240-01-312-6281
NSN 5925-01-463-9258
NSN 5998-00-766-7432
NSN 5925-00-920-7066
NSN 5920-01-537-5541
NSN 2925-01-534-5389
NSN 6650-01-155-8182
NSN 5930-01-648-2582
NSN 5930-01-227-4385