Product Features
C/o two test wells, each well is 2.6"dia x 11/3" deep (60 cubic inches) and have a 0.25" deep mercury trap. Transfer fluids: cold bath - halocarbon 0.8; hot bath - dow 710.
Detail Specifications
Overall Depth:
26.250 inches
Overall Height:
45.000 inches
Overall Width:
45.250 inches
Ac Voltage Rating:
Between 208.0 volts and 230.0 volts
Frequency Rating:
60.0 hertz
Mounting Configuration:
Free standing cabinet on castors
Unpackaged Unit Weight:
450.000 pounds
Calibration Type:
Deg celsius and degrees fahrenheit
Instrument Calibration Range:
Cold bath -100 to +70 degrees c (-148 to +158 degrees f); hot bath +60 to +300 degrees c (+140 to +572 degrees f)
Calibration Accuracy Rating:
100 to -40 degrees c - porm 5 degrees c; -40 to +150 degrees c - porm 1 degrees c; +150 to +300 degrees c - porm 5 degrees c
Functional Description:
Consists of 2 independently operable bath systems. Each bath has its own tank with a constant level test well & pump. Each has its own temperature controller, probe, & heating and colling systems. Seperate front panel power controls, labeled hot bath and cold bath, allow independent power up and use.
Reference Data And Literature:
Commercial manual
Item Name:
Calibrator,thermometer
Manufacturer Part Numbers
Part Number
Cage
RNCC
RNVC
RNAAC
Status
MSDS
SADC