Battery operated; capable of resistive ratio measurements with 0.1 parts per million resolution over the normalized ratio range of zero to one; portable; accuracy measurements of 0.01%; housed in a rugged enclosure; includes hinged instrument cover which is a compartment for storage of the mach trim actuator interface cable p/n 1683-2600-11, extra fuses and the tech manual; null detector uses 5 leds for operator feedback in place of an analog meter; sensitivity from +/- 1 microvolt to +/-50 microvolt from a front panel control; power consumption 280 mw; operating temp range -10 degrees c to +55 degrees c
A test instrument which will measure resistance and voltage ratios by the direct ratio method of measurement and will also measure resistance and voltage values.
Classification: Apparatus to test voltage, current or resistance without a recording device
NSN 5915-00-670-7948
NSN 6620-01-428-3930
NSN 6145-01-245-8856
NSN 5999-01-319-6854
NSN 6105-00-777-8139
NSN 5835-01-381-4452
NSN 2925-00-678-5502
NSN 6650-01-013-0667
NSN 4140-01-644-9823
NSN 6220-01-581-4801
NSN 5915-01-571-7341
NSN 5855-01-045-9728