Item shall perform withing following climatic conditions: temperature, non-operating, -46 degrees c to +70 degrees c/operating, 0 degrees c to 55 degrees c; operating characteristics shall be as follows: blood pressure heart rate shall be 30-240 bpm, bp outputs shall be 0-200 mm hg +/- 2%, unit have selectable 120/80 hg +/- output, static and dynamic wave forms, ecg simulator shall be variable, respiration rate be variable, temperature output be selectable; unit have electrode lead continuity test feature; selector switc (s) /terminals describe function, range, test; unit furnished with protective case