Shall be a halogen light source with a screw-on faceplate, a water bottle assembly and a built-in air pump; for use with fiberoptic endoscopic procedures requiring insufflation. Shall use a 150 watt halogen lamp lasting approx 50 hrs; voltage requirements shall be 110/220, 50/60 hz, with forced air cooling; shall have a vibrator type air pump with 0.35kg/cm2 pump pressure; shall be supplied with a screw-on faceplate receptacle to allow use of multiple manufacturers style of flexible scopes; shall come supplied with a water bottle assembly for irrigation. Bottle shall contain 500cc water and be easy to fill and clean; shall have an on/off switch for both lamp and air pump; switches shall be labeled; shall have handle for carrying purposes; light source shall be that normally furnished by the manufacturer as commercial practice; metal portions shall be manufactured of a suitable corrosion-resistant metal alloy that is properly hardened & passivated