Portable liquid oxygen storage and distribution system; provides life support oxygen gas while on board an aircraft in flight; npt lox unit fills from any usaf or nato cru-50/a fill connector; capacity: 20 l lox; service: up to 6 patients; delivery rate: up to 66 lpm at 50 +/- 5 psig; operating temperature: 32 to 120 degrees f; storage temp: -40 to +158 degrees f; altitude: sea level to 15000 feet; relief valve settings, system: 100-120 psig, safety: 170-200 psig; rupture disc burst pressure: 200-250 psig at 72 degrees f; optimal time to performance after filling: 10 minutes; weight (full): 148 lbs
NSN 5920-01-425-2336
NSN 5965-00-791-0964
NSN 5915-01-494-7452
NSN 5990-01-468-1292
NSN 6620-00-852-8245
NSN 5945-01-040-1655
NSN 5915-01-068-4441
NSN 5965-01-150-6315
NSN 6650-01-148-3499
NSN 7035-00-517-8232
NSN 6240-01-411-7482
NSN 6220-01-418-1649