Shall be capable of connecting toand testing movement thermo-couples; meter-movement resistance temperature detectors (rtds); servo thermo-couples ; and commercial thermo-couples type k/t/j and tf-30, case shall be no more than 20 inches long, 15 inches wide, and 15 inches high, case will have a handle attached and a lockingmechanism p rovided, weight shall be no more than 35 pounds; interface shall be providedthat will display values being measured or sourced while operating between 50 and 100 degrees farenheit, temperature calibrator shall have an on/off power switch; battery charge indicator; source/measure switch; output voltage tuning ability from 0-35volts down to .5mv of the desired output; mode select capability; ability to test single or double rtd indicators; ability to source 22.5 or 28.5 volts plus or minus .5mv; ability to simulate rtd and synchro indicator voltage ranges, will include 115/230v AC at 60hz power cable
NSN 6135-01-551-0562
NSN 6105-01-202-8706
NSN 5950-00-468-2209
NSN 6110-01-197-6954
NSN 5920-01-599-1559
NSN 5930-01-212-1696
NSN 5960-00-027-3681
NSN 7035-01-486-4118
NSN 6240-01-047-2786
NSN 6130-01-492-5841
NSN 6150-01-088-7643
NSN 5998-01-067-8494
Last Updated: 12/28/2024